油扩散泵 DIP 30.000

抽速 (N₂): 30000 l/s
排气口: DN 160 ISO-K
进气口: DN 800 ISO-F
登录后查看价格

电机:

节能控制单元:

Gesamte Liefermenge:
in  1  Lieferungen

油扩散泵 DIP 30.000

The pumps in the DIP series are high-vacuum pumps (<10-2 mbar). They are always operated in conjunction with forevacuum pumps. The DIP series pumps are water cooled and utilize the oil diffusion principle in their operation. They are employed in high-vacuum technology to evacuate vacuum chambers. They achieve their highest pumping speeds in pressure ranges from 10-2 to 10-7 mbar.

The diffusion pumps in the DIP series comprise the following component assemblies:

  • Water-cooled pump housing with high-vacuum, and forevacuum connection flanges
  • Nozzle assembly
  • Vaporization chamber with heating elements
  • Cold cap baffle
  • Forevacuum baffle
  • Energy Efficiency Control or electrical junction box

 

Nozzle system

The DIP pumps are fitted with a four-stage nozzle system made of light-alloy metal and with an internal heating system comprising heating cartridges and mounting wells to which heat diffusion fins are soldered. The mounting wells are made of stainless steel and are welded vacuum-tight in the pump body, in a horizontal position.

The housing for the DIP pump is made of standard grade steel; the high-vacuum connection flange and the forevacuum connection are made of stainless steel (alloy 1.4301), the cooling coils of copper, and the cold cap baffle of nickel-plated copper.

 

Heat diffusion fins

The heat diffusion fins are made of copper and are only partially immersed in the pump fluid in the vaporization chamber.

The section of the heat diffusion fins immersed in the pump fluid is dimensioned so that that intense but surge-free vaporization of the pump fluid is achieved.

The sections of the heat dissipation fins located above the level of the pump fluid apply additional energy to the pumping vapor.

 

Thermostat

To protect the heating element, a thermostat sensor is attached to a heat diffusion fin which protrudes from the fluid; this will switch off the pump’s heaters as soon as the temperature set at the safety thermostat is exceeded.

The sensor has to be installed into the system control by the customer so that it will safely switch off the power.

 

Heating cartridges

The heating cartridges can be easily replaced when required. It is not necessary to dismantle the pump to do so.

 

Cold cap baffle

To prevent fluid from flowing back into the vacuum vessel, the DIP series pumps are fitted with a water-cooled cold cap baffle in the area of the intake port.

A water-cooled forevacuum baffle located on the forevacuum side effectively prevents fluid being swept into the forevacuum unit.

 

Technical Data
DIP 30,000
High-vacuum connection DN 800 ISO-F
Forevacuum connection DN 160 ISO-K
Pumping speed for
hydrogen <10-4 mbar
air1) <1 · 10-4 mbar

l · s-1
l · s-1

30,000
24,500
Working range mbar <10-2 - <10-7
Ultimate total pressure 2) mbar <5 · 10-7
Max. permissible forevacuum pressure mbar 6 · 10-2
Pump fluid fill, min./max. l 10 / 15
Line power supply 5)
Standard, 50/60 Hz
Special, 50/60 Hz

V
V

(± 10 %) 230/400 ~ 3 Ph Δ/Y
460 ~ 3 Ph Δ
Heating power kW 18
Heating cartridges   18
Warm-up period min <30
Coolant (minimum)
for the pump 3)
for the cold cap baffle

l · h-1
l · h-1

900
80
Number of cooling circuits
(including cold cap baffle)
 
3
Coolant connection
Pump
Cold cap baffle

G
G

1/2"
3/8"
Weight, approx. kg 296
Recom. forevacuum pumps 4)
at working pressures >10-4
at working pressures <10-4

mbar
mbar

SV 300 + W 1001
SV 300 + W 251

 

1) Measured as per DIN 28 427 using LVO 500 normal as the pump fluid.
2) Measured as per DIN 28 427 using LVO 500 normal as the pump fluid. When using the LVO 520 pump fluid and FPM (fluoroelastomer) gaskets, the DIP pumps with water-cooled baffles will achieve pressures below 1·10-7 mbar following suitable bake-out procedures.
3) The coolant water volume is referenced to ΔT = 10 K. The discharge temperature should not exceed 50 °C.
4) Single- and two-stage rotary vane pumps (TRIVAC; SV) from our line of forevacuum pumps in conjunction with roots pumps (RUVAC) in pumping systems.
5) The EEC is connected both with a load voltage, and a control voltage (of 24V ±15%, 50/60Hz, 50VA, 6A or 230V ±15 % 1 ph). Both voltages will be disconnected safely from the mains via the EPO / main switch.


类似产品

Welcome to Leybold Welcome to Leybold

請選擇你的國家 Please select your country.

Ihr Land ist nicht dabei? Besuchen Sie unseren globalen Shop. Your country is not listed? Visit our global shop.

在线留言
亲爱的用户,

我们总是期望能够为您提供最好的服务和最优的品质,但是,却又不能阻止产品相关问题的发生。为了帮助您解决产品的相关问题,我们制作了一份申请表,它可以使您以最便捷的方式联系到我们, 如需获得申请表,请使用如下链接.

如果遇到关于我们网上商店的一些常见问题, 请阅读

敬上
您的莱宝团队

请留下您的需求和联系信息